Memo-Graph S 9 Trouble-shooting
Endress+Hauser 81
9.4 Spare parts
!
Note!
Installation instructions are included with the spare part.
9.4.1 Spare parts diagram
Figure 8: Spare parts
Modem transmission does not func-
tion
Modem at Memo-Graph S has not been initialised Initialise modem with PC software
Wrong cable between modem and Memo-Graph S Use original cable RSG12A-S2
Device address or release code of Memo-Graph S and PC
software do not match
Set identical device address and release code in
Memo-Graph S and PC software
Etrhernet connection does not func-
tion
IP address, subnet mask or gateway are not set correctly Check settings and correct them
Interface does
not function
Cable defective
Incorrect connection assignment
Incorrect address
Incorrect interface parameters
Replace cable (Accessories see Chapter 8)
Please use original cable
Check and set correctly.
Check and set correctly.
No data on ATA
flash card
Set-up change
Software update / upgrade
ATA flash card defective
ATA flash drive defective
Save the data onto a data carrier before any changes to
the set-up.
Save measured values onto a data carrier before any
changes to the software.
Replace ATA flash card
Replace ATA flash drive
Problem Cause Remedy